VACUUM ›› 2019, Vol. 56 ›› Issue (1): 1-5.doi: 10.13385/j.cnki.vacuum.2019.01.01
ZHAO Yue-shuai, SUN Li-chen, SHAO Rong-ping, YAN Rong-xin, SUN Wei, LI Zheng
Received:
Online:
Published:
Abstract: Four refrigerator cooled cryopumps (DN1250 LN2) were designed and manufactured by BISEE. The pumps were used in space environment simulation system to meet the demand of the high vacuum in thermal vacuum test. The design and test method for pumping speed, cool down time and crossover were introduced in this paper. A test system was built to test the main performance of the cryopumps. The experimental results show that the pumping speed for N2 of the pump is up to 57000 L/s, the cool down time is about 330min, and the crossover is over than 3.0 ×105Pa·L.
Key words: cryopump, pumping speed, cool down time, crossover, test method
CLC Number:
ZHAO Yue-shuai, SUN Li-chen, SHAO Rong-ping, YAN Rong-xin, SUN Wei, LI Zheng. Design and performance test of DN1250 LN2 refrigerator cooled cryopumps[J].VACUUM, 2019, 56(1): 1-5.